Product Overview
Features
  • Ranges from 200 mbar to 700 bar
  • Total accuracy to ±0.1 % FS
  • Stainless steel construction
  • Wide temperature range -40 °C to 125 °C
  • Fast update rate 1 ms
  • Customer set filter setting
  • Baud rate upto 1000k bit/s
  • Excellent long-term stability
Specifications

Operating pressure ranges

Gauge ranges
Any range from 200 mbar to 100 bar (3 to 1500 psi)

Ranges up 70 bar are enabled in negative gauge to the
range or a maximum of -1 bar (-15 psi)

Absolute ranges
Any range from 700 mbar to 700 bar (10 to 10000 psi)
All ranges are zero based

Wet Dry differential
Any range from 200 mbar to 35 bar (3 to 500 psi)
All units are bidirectional
Maximum line pressure: 70 bar

Wet Wet differential
Any range from 700 mbar to 35 bar (10 to 500 psi)
All units are bidirectional
Maximum line pressure: 70 bar

Ordering
When ordering, specify maximum working pressure.
The lower pressure limit will be specified inline with rules
stated above.

Over pressure
At least 2 × Pressure range with negligible calibration change

For differential versions, the negative side must not exceed its
positive side by more than:
2 × Pressure range up to a maximum of 15 bar

Containment pressure
4 × Pressure range for gauge ranges
(200 bar (3000 psi) maximum)
200 bar (3000 psi) for absolute ranges below 100 bar
(1500 psi)
1200 bar (17000 psi) for absolute ranges above 100 bar
(1500 psi)
For differential versions, the negative side must not exceed its
positive side by more than:
2 × Pressure range up to a maximum of 15 bar

Supply
Voltage: 5 Vdc to 32 Vdc
Current: <30 mA

Output/communications
CANopen V2.0B

CAN setting options (select when ordering)
Node ID: Can be set to any integer between 2 and 127
(default setting is 2)
Baud rate: Can be set to 10k, 20k, 50k, 125k, 250k, 500k, 800k,
1000k (default setting is 250k)

Power on time
500 ms to acquisition from power on

Insulation resistance
>100 MΩ at 500 Vdc performance
 

Performance

Pressure performance
Accuracy over the calibrated temperature range including
zero and span setting and the effects of non-linearity,
hysteresis and repeatability

Gauge/absolute/wet dry differential
A3 Premium: ±0.1 % FS over -20 to 80 °C (-4 to 176 °F)
A2 Improved: ±0.2 % FS over -40 to 125 °C (-40 to 257 °F)

Wet wet differential
Premium: ±0.2 % FS over -20 to 80 °C (-4 to 176 °F)
Improved: ±0.4 % FS over -40 to 125 °C (-40 to 257 °F)
Increases pro-rata for pressures below 700 mbar

Temperature performance
Accuracy over the calibrated temperature range
±3 °C (6 °F) with 0.1 °C (0.2 °F) Resolution

Long term stability
±0.05 % FS/year typical
±0.1 % FS maximum at reference conditions
Increases pro-rata below 700 mbar (10 psi)

Line pressure effects (differential sensors only)
Zero shift: <±0.03% span/bar
Span shift: <±0.03% span/bar
Effects increase pro-rata for ranges below 700 mbar

Physical specifications

Environmental protection
IP67

Operating temperature range
-40 °C to 125 °C (-40 °F to 257 °F)

Pressure media
Fluids compatible with stainless steel 316L and Hastelloy C276
Statement in accordance with the European pressure
equipment directive

Enclosure materials
Stainless steel 316L (body)

Pressure connectors

  • PA G1/4 female
  • PB G1/4 male flat
  • PE 1/4 NPT female
  • PF 1/4 NPT male
  • PJ M14×1.5 60° Int Cone
  • PZ M10×1 80° Int Cone

Other connectors may be available. Contact Druck to discuss
your requirement.