Druck Knowledge Base | FAQ | Pressure Stations

Find below Drucks knowledge base and FAQ for Pressure Stations

If you'd like to contact Druck directly, please click here or email us at druckcontact@bakerhughes.com 

For all other FAQ content click here.

FAQ | Pressure Stations


FAQ | Pressure Stations
What is the internal volume of the PV621 pneumatic assembly?

The internal volume of the pneumatic assembly is approximately 1cc when the Volume Adjuster is fully in (clockwise), when the Volume Adjuster is fully out (anti-clockwise) the volume is approximately 9cc.
 

What is the internal volume of the PV622 pneumatic assembly?

The internal volume of the pneumatic assembly is approximately 1cc when the Volume Adjuster is fully in (clockwise), when the Volume Adjuster is fully out (anti-clockwise) the volume is approximately 14cc. 
 

Can the PV624 be used with a pressure module higher than 20 bar?

The PV624 will work with a PM620 or a PM620T with a greater than 20bar full scale but the set point entry and therefore the maximum range of a PV624 is limited to 20bar/300psi.
 

Can PV624 / DPI620G purchased without Bluetooth have the option fitted at a later date?

A chargeable service to upgrade to a Bluetooth version is available as long as we have a Bluetooth licence for the Country of use.
The unit will have to be returned to Groby for this.
The DPI 620G will have a maximum age limit as some really old units cannot be done.
 