Three model choices to generate from 95% vacuum to 1000 bar/ 15000 psi
- 20 bar/300 psi in 30 seconds
- 100 bar/1500 psi without gas bottles and regulators
- 1000 bar/15000 psi hydraulic pressure
DPI 612 pFlex
General Specifications:
- Maximum working pressure 20 bar / 30 psi pneumatic
- Maximum vacuum -95% of atmospheric pressure
- Pressure media Non-corrosive gases
- Pressure connection Quick-to-fit connection system. G1/8 female and 1/8 NPT female adaptors are provided
- Size and weight 350 mm x 170 mm x 145 mm ( 13.7 in x 6.7 in x 5.7 in) excluding accessories 3.4 kg (7.5 lb) including battery and pressure module
- Generation system Pressure/vacuum selector, high-efficiency pump, fine adjuster, precision release valve
DPI 612 pFlexPro
General Specifications:
- Maximum working pressure 100 bar / 1500 psi pneumatic
- Maximum vacuum -95% of atmospheric pressure
- Pressure media Non-corrosive gases
- Pressure connection Quick-to-fit connection system. G1/8 female and 1/8 NPT female adaptors are provided
- Size and weight 350 mm x 170 mm x 160 mm ( 13.7 in x 6.7 in x 6.3 in) excluding accessories 4.1 kg (9 lb) including battery and pressure module
- Generation system Pressure/vacuum selector, high-efficiency pump, pressure intensifier/fine adjuster, non-return valve, precision release valve
DPI 612 hFlexPro
General Specifications:
- Maximum working pressure 1000 bar / 15000 psi hydraulic
- Pressure media De-mineralized water or mineral oil (ISO viscosity grade < 22). 100cc internal reservoir
- Pressure connection Quick-to-fit connection system with G1/8 female and 1/8 NPT female adaptors
- Size and weight 350 mm x 170 mm x 170 mm ( 13.7 in x 6.7 in x 6.7 in) excluding accessories 4.7 kg (10.4 lb) including battery and pressure module
- Generation system Priming pump, pressure intensifier/fine adjuster, non-return valve, precision release valve
Interchangeable Pressure Modules
The PM 620 is the latest development in digital output sensor technology incorporating a number of key innovations to allow pressure re-ranging of compatible equipment. A simple screw fit makes both the pressure and electrical connections without the need for tools, sealing tape, cables, or plugs and digital characterization allows interchangeability without set-up or calibration. The PM620 also comes in an intrinsically safe version, the PM620S.
PM620 now incorporates our unique range of TERPS resonant silicon pressure sensor technology, the PM620T and the PM620TS intrinsically safe version. Providing up to four times greater stability and higher accuracy than current pressure measurement technologies.
Note – accessories identified by * are supplied as standard with applicable DPI 612 models except DPI 612-XXX–00.
*Carrying case (P/N IO620-CASE-3) A protective carrying case with shoulder strap and a large pocket for accessories.
Spare/replacement rechargeable battery (P/N CC3800GE) Spare/replacement lithium-ion battery for the DPI611 and DPI 612 Flex series.
Spare/replacement mains adaptor (P/N IO620-PSU) A universal input mains adaptor connects to the DPI 612 to charge the battery and power the instrument. Input voltage 100 to 240 VAC 50/60 Hz. Mains socket adaptors are provided.
Battery charging station (P/N CX6100GE) This external battery charging station and universal input mains adaptor allows a spare battery to be charged independently of the DPI 612 for minimum instrument downtime. A complete charge cycle takes approximately 6.5 hours. (CX6100GE Not available in Korea).
USB cable (P/N IO620-USB-PC) Connects the DPI 612 to a PC.
IDOS to USB converter (P/N IO620-IDOSUSB) Allows connection of an IDOS universal pressure module to the DPI 612. P/N IO620- USB-PC is also required to connect the converter to the DPI 612 USB port.
USB to RS 232 cable (P/N IO620-USB-RS232) Connects the DPI 612 to an RS 232 interface
*Dirt and moisture trap Prevents contamination of the DPI 612 pFlex pneumatic systems and cross-contamination from one device under test to another. The trap connects directly to the DPI 612 pFlex pressure port and replicates the quick-fit connection for compatibility with the hose and adaptor kits. Not for use with DPI 612 hFlex models.
- *P/N IO620-IDT621: Maximum working pressure 20 bar / 300 psi.
- *P/N IO620-IDT622: Maximum working pressure 100 bar / 1500 psi.
Quick-to-fit hose and adaptor system These high-pressure hoses and adaptors are compatible with the pressure ports of the DPI 612 Flex, DPI 611, and DPI 620 series and are quick-to-fit without tools or sealing.
- *IOHOSE-NP1: Pneumatic hose 1m/3.28ft. Maximum pressure 20 bar/300 psi.
- *IOHOSE-NP2: Pneumatic hose 2m/6.56ft. Maximum pressure 20 bar/300 psi.
- *P/N IO620-HOSE-P1: Pneumatic hose 1m/3.28ft. Maximum pressure 400 bar/5800 psi.
- P/N IO620-HOSE-P2: Pneumatic hose 2m/6.56ft. Maximum pressure 400 bar/5800 psi.
- *P/N IO620-HOSE-H1: Hydraulic hose 1 metre/3.28 ft. Maximum pressure 1000 bar/15000 psi.
- P/N IO620-HOSE-H2: Hydraulic hose 2 metre/6.56 ft. Maximum pressure 1000 bar/15000 psi.
- P/N IO620-BSP: Pressure adaptors G1/8 male and G1/4 male, G1/4 female, G3/8 female, and G1/2 female.
- P/N IO620-NPT: Pressure adaptors 1/8” male and 1/4” male, 1/4” female, 3/8” female, and 1/2” female.
- P/N IO620-MET: 14 mm female and 20 mm female.
P/N IO620-COMP Comparator adaptor For greater efficiency, two test devices can be connected at the same time. The adaptor connects to the pressure port of the DPI 612 and provides two outlet ports.
*Pressure relief valve Protects the PM 620 pressure module and the device under test from overpressure.
Screen protector (P/N IO61X-PROT)
Replacement touch screen stylus (P/N IO61XStylus)
Replacement test lead set (P/N IO61X-Lead)
Replacement shoulder strap (P/N IO612-STRAP)
Datasheets - DPI 612